| Sign In | Join Free | My ecer.co.uk |
|
Brand Name : Haida
Model Number : HD-E802
Place of Origin : Guangdong, China (Mainland)
MOQ : 1 Set
Price : Negotiable
Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability : 30 Sets per Month Haida
Delivery Time : 15 days after order confirmed
Packaging Details : Export wooden case
Chamber dimension (W*D*H) : 1300*500*1460mm
Chamber material : SUS#304 stainless steel
Temperature range : RT to 70℃
Temperature fluctuation : ≤±1.0℃
UV Light Accelerated Aging Environmental Test Chamber
Generl Specifications
Packaging Dimensions: (WxDxH) 1490*620*1620mm
Power supply source: single-phase, 220V±10%, 50Hz (can be appointed)
Gross Weight: 180kg
HS Code: 8479899990
Standard Features
| Item | Specification |
| Chamber dimension (W*D*H) | 1300*500*1460mm |
| Chamber material | SUS#304 stainless steel |
| Temperature range | RT to 70℃ |
| Temperature fluctuation | ≤±1.0℃ |
| Temperature uniformity | 3℃ |
| Humidity range: | ≥75%RH |
| Controller | Programmable controller, LCD touch screen |
| Control mode | Balance temperature humidity control (BTHC) |
| Test cycle setting | Exposure, condensation and water spray test cycle is programmable |
| Water spray cycle | Spray 18mins in every 120mins / spray 12mins in every 60mins |
| Water usage | 8L/day |
| Distance from sample to lamp | 55±2mm |
| Centre distance betweenlamps | 65~70mm |
| Irradiance range | 0.45~0.90W/m2 |
| UV lamps | Imported Atlas UV-A: 315-400nm (8pcs, 1600h lifetime) |
| Lamp power | 40W/Piece |
| Specimen Size | 75×290mm (24pcs) / 75x150mm (48pcs), maximum thickness 5mm |
| Testing time | 0~999H, adjustable |
| Protection system | Overload short circuit protection |
Applications
Test Criteria
ISO 4892-3
ASTM G154

|
|
UV Light Accelerated Aging Environmental Test Chamber Images |